BASIC Step-by-Step Instructions for JEOL 6060LV
Version 4/13/11
General Operation
1. Check System: Pumps running and EVAC light green (main panel near key switch)?
If not notify a staff member.
2. If the SEM software is not running, start SEM Software (SEM Main Menu icon). It
takes several minutes to start, be patient. If you need to log into computer use the
account (USER:jeol;PASSWORD:jeol). Please do not log into other accounts. Use
portable media, ftp, or UIUC NetFiles to transfer your data.
3. Prepare your samples before venting! Please keep the area clean, organized, and wear
GLOVES!! Anything that goes in the vacuum should be clean and handled only with
gloves or tweezers; keep the vacuum clean.
4. Press the Sample button on toolbar.
5. Press the Vent button; it takes a couple minutes to vent.
6. When Vent light stops blinking (main panel near key switch), slide open stage
drawer.
Page view 0
1 2 3 4 5 6 ... 13 14

Summary of Contents

Page 1 - Version 4/13/11

BASIC Step-by-Step Instructions for JEOL 6060LV Version 4/13/11 General Operation 1. Check System: Pumps running and EVAC light green (main panel n

Page 2

5. If the image is not sharp enough, try decreasing the spot size, and verify the sample is at a working distance of 8-10mm. Also lower kV operation

Page 3

c. SHADOW (a+b+c) – image with strong oblique component, image similar to SEI but with BSE, needed for LV mode where SEI detector does not function T

Page 4

Changing the Vacuum Mode LV mode is generally used to observe and analyze non-conducting sample without the need for a conductive coating (Au/Pd or C

Page 5

Troubleshooting If the SEM software is not responding or is behaving abnormally you can close the application and restart it (PC-SEM). If this does

Page 6

JEOL 6060LV GUI Menu bar When each menu is clicked, a pull-down menu is indicated. If you click a pull-down menu, the various operation and setup

Page 7

7. Slide sample holder onto stage dovetail. Note proper direction to slide on “dovetail” relative to step on stage platform. 8. Adjust sample hei

Page 8

9. Specimen tilting is limited by working distance and specimen diameter. The IR chamber-scope must be used for sample tilting and Z adjustment to

Page 9

15. Set the Signal to SEI , if not already (You can change it to BEIW, if desired later). 16. Click the HT Ready icon to turn on the High Voltage,

Page 10

e. Manually check filament saturation. This is best done at a higher spot size setting (40-60) and a high contrast setting. Click the Filament Heati

Page 11

running auto alignment first select an area on the sample or sample holder that has homogeneous brightness at moderate magnification (1-5 kx) to get t

Page 12

c. Hint: For easier operation, in HV mode, use the SEI detector to set up the instrument and focus, etc. Switch to the BEIW mode mostly to look at or

Page 13

Changing Samples or Ending Your Session 1. Turn off the High Voltage (click HT On control on main panel, it should change to HT ready) 2. Wait 2

Page 14

Aligning the Objective Aperture 1. This procedure needs done if the image shifts in any direction when going through focus. 2. Center a distinct fe